Research AbstractEMS World Expo · Scientific Symposium217 First Responders
Short Summary
This study involved 217 first responders from fire departments in Tennessee. Participants performed 60 seconds of manual ventilation on a simulated lung without the Sotair device. After a brief educational intervention on using the Sotair flow-rate limiting device, they repeated the 60-second ventilation session with Sotair attached to the manual resuscitator bag. The study compared peak inspiratory pressures, tidal volumes, and respiratory rates before and after the intervention.
Key Findings
- Lower Peak Inspiratory Pressures: Sotair reduced peak pressures (15.7 cmH₂O vs. 17.6 cmH₂O, p < 0.01).
- More Controlled Tidal Volumes: Sotair provided lower tidal volumes (525 mL vs. 594 mL, p < 0.01).
- Improved Ventilation Consistency: After a brief educational intervention, providers ventilated more consistently with Sotair, demonstrating its ease of adoption and effectiveness in real-world emergency settings.

Conclusion
Reference
Kumar P, Holley JE, Justice JM, Slutsky AS, Brady MF. Manual ventilation performance in first responders using a flow-rate limiting device (Sotair). Presented at: EMS World Expo’s International Scientific Symposium in partnership with UCLA’s Prehospital Care Research Forum; September 2024; Las Vegas, NV.